Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
Investigation results: root cause: other, low water flow.- insert tip has low water flow.Also insert was tested on a digital thermometer i.D.# (b)(4) due date: (b)(6) 2023.The temperature was 102.1° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Conclusion code: unexpected failure.Returned qty.(b)(4) insert, (b)(4), ga, warranty, test method / gp005-wi02.Inductance: gauge id# (b)(4) , due: (b)(6) 2024, result: 3.19.Meets spec, does not meet spec, n/a.Tip condition: - meets spec does not meet spec n/a stack condition, meets spec, does not meet spec, n/a.Tested on: g136 gage id# (b)(4), due date: (b)(6) 2023, set pressure: 35 psi.Water flow: output rate: (b)(4)- meets spec, does not meet spec, n/a.Spray pattern: - meets spec , does not meet spec, n/a.Water leak: hp sealing ring, proximal ring, distal ring, seam leak , n/a.Vibration: - meets spec, does not meet spec, n/a.Grip condition: meets spec, does not meet spec , n/a.Other visual observation: insert tip has low water flow.Also insert was tested on a digital thermometer i.D.# (b)(4) due date: (b)(6) 2023.The temperature was 102.1° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed - not confirmed.
|