Model Number BUTTON PROCESSOR 3 CONTROL UNIT [ME1550]
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Event Description
The user's device reportedly melted around the magnet area during the charging process.Further information has been requested.
 
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow-up report.
 
Manufacturer Narrative
Additional information: according to the information received from the field the rondo 3 audio processor was melting at the magnet site whilst charging.It is assumed that the audio processor was placed with the wrong site on the charger.However as the device has not been received for investigation a root cause for the reported issue cannot be determined.
 
Event Description
The user's device reportedly melted around the magnet area during the charging process.The user was not harmed in any way by the heating of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16272380
MDR Text Key308501186
Report Number9710014-2023-00089
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09008737392468
UDI-Public(01)09008737392468
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberBUTTON PROCESSOR 3 CONTROL UNIT [ME1550]
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received01/31/2023
Supplement Dates Manufacturer Received01/23/2023
Supplement Dates FDA Received05/25/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexFemale
-
-