• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71733-01
Device Problems Device Alarm System (1012); No Audible Alarm (1019)
Patient Problems Hypoglycemia (1912); Chills (2191); Loss of consciousness (2418)
Event Date 01/20/2023
Event Type  Injury  
Event Description
An alarm issue was reported with the adc device.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced hypoglycemia, cold(temperature) and a loss of consciousness.Customer was unable to self-treat requiring treatment of provision of sugar and water by the by third-party.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
Adc has identified a software defect for the freestyle libre 2 application with the android 13 os where the application may experience intermittent signal loss.As a result, the application user may not receive glucose results and/or glucose alarms and may not be alerted of low or high glucose conditions.Based on the investigation, this complaint is confirmed.This issue was addressed in the field by adc fa1010-2023.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the adc application in use with a samsung a12 13-2819305 with android 13 operating system.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced hypoglycemia, cold(temperature) and a loss of consciousness.Customer was unable to self-treat requiring treatment of provision of sugar and water by the by third-party.There was no report of death or permanent injury associated with this event.The impacted product associated with this complaint is on market in the united states as well as the following countries ous: au, at, be, hr, cz, dk, fi, fr, de, gr, il, it, kw, lu, nl, nz, no, pl, pt, qa, sa, es, se, ch, tw, ae, gb, ca.Field action fa1010-2023 was issued to all impacted countries 08feb23.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Type of Device
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key16272902
MDR Text Key308485402
Report Number2954323-2023-04811
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71733-01
Device Catalogue Number71733-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/21/2023
Initial Date FDA Received01/31/2023
Supplement Dates Manufacturer Received03/07/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-