MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q180V
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2021
Event Type  malfunction  
Manufacturer Narrative
This mdr is being submitted as part of a retrospective review and remediation effort based on enhancements made to the company¿s mdr and complaint handling processes.Capas have been opened to manage the actions that are being taken to remediate this issue and ensure any required mdr reporting is completed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 10 years since the subject device was manufactured.Based on the results of the legal manufacturers investigation, although a conclusive root cause could not be determined, it is likely the forceps could not be inserted or removed due to following: the event occurred due to deformation of the forceps channel.The event occurred due to breakage of the instrument.The event occurred due to foreign material in the forceps channel.The following information is stated in the ifu (instructions for use) which may help to prevent the issue: operation manual - 3.8 inspection of the endoscopic system, inspection of the instrument channel, and forceps elevator.Operation manual; important information ¿ please read before use: precautions.Operation manual - 3.6 inspection of ancillary equipment.Operation manual - 4.3 using endotherapy accessories.Reprocessing manual - 5 reprocessing the endoscope (and related reprocessing accessories).Reprocessing manual - 7 reprocessing endoscopes and accessories using an aer/wd.Olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported that there was suction on the channel and something was stuck.The cleaning brush could not pass through.The reported issue was observed during an unspecified procedure.There was no report of patient or user injury due to the event.The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, it was observed that the forceps could not be inserted or removed.This report is being submitted for the reportable malfunction of forceps could not be inserted or removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16274684
MDR Text Key309310635
Report Number9610595-2023-01573
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170339967
UDI-Public04953170339967
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/05/2021
Initial Date FDA Received01/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0379-2022
Patient Sequence Number1
-
-