Model Number 1035-48-000
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Joint Dislocation (2374); Joint Laxity (4526); Physical Asymmetry (4573)
Event Date 01/22/2023
Event Type  Injury  
Event Description
Patient was revised due to the hip unfortunately dislocated.The surgeon tried for a closed reduction and was unsuccessful.He felt there may be some soft tissues in the joint space that was causing instability of the bipolar.When opening the hip, tissue impingement was confirmed.He felt the patient may have been a little short as well, so he trialed a 48 bipolar head with a 28 +8.5 head and found the hip to be more stable.The real bipolar and 28 head was opened and implanted.The hip was reduced and stability was confirmed.The closing process began.There was no surgical delay.Doi: (b)(6) 2023, dor: (b)(6) 2023, affected side: right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 48X28 BRN
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16277863
MDR Text Key308553281
Report Number1818910-2023-02608
Device Sequence Number1
Product Code KWY
UDI-Device Identifier10603295003380
UDI-Public10603295003380
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1035-48-000
Device Catalogue Number103548000
Device Lot NumberM1146K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ARTICUL/EZE BALL 28 +1.5 GR
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexMale
-
-