BECTON, DICKINSON AND COMPANY (BD) BD PRESET¿ ECLIPSE¿; BLOOD SPECIMEN COLLECTION DEVICE
|
Back to Search Results |
|
Catalog Number 364391 |
Device Problem
Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during use with bd preset¿ eclipse¿ the plunger was discovered to be deformed.There was no report of patient impact.The following information was provided by the initial reporter, translated from japanese to english: during blood collection, the hcp felt that it was a bit difficult to move the plunger, and then noticed that the rubber stopper seemed to be melting.
|
|
Manufacturer Narrative
|
Investigation summary: bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and the indicated failure mode for deformed plunger stopper was observed.Additionally, 10 retention samples from bd inventory were evaluated by visual examination and upon completion the indicated failure mode for deformed plunger stopper was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode deformed plunger stopper.Bd has initiated further root cause investigation relating to the issue of deformed plunger stopper through corrective and preventive actions.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|