• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG INVENTRA 7 HF-T QP DF4 IS4 PROMRI; CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG INVENTRA 7 HF-T QP DF4 IS4 PROMRI; CRT-D Back to Search Results
Model Number 393011
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Event Description
After an implantation period of approx.76 months, it was observed via homemonitoring that the device shows the eos status.The device was explanted.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the icd was subjected to an electrical analysis, revealing that the device could not be interrogated.Therefore, the device was opened, and the inner assembly was inspected.The visual inspection of the inner assembly showed no anomalies.Subsequently the current consumption of the electronic module was verified by direct measurement and proved to be as expected.There was no indication of a malfunction of the electronic module.Battery voltage measurement confirmed a depleted battery, which could be attributed to an increased internal self-depletion within the battery.Please note that this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVENTRA 7 HF-T QP DF4 IS4 PROMRI
Type of Device
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16278659
MDR Text Key308565436
Report Number1028232-2023-00532
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Model Number393011
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/01/2023
Initial Date FDA Received02/01/2023
Supplement Dates Manufacturer Received09/11/2023
Supplement Dates FDA Received09/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-