• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. OXINIUM FEM HD 12/14 36 MM M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71343604
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Foreign Body In Patient (2687)
Event Date 08/13/2020
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case(b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of the national joint registry of england, wales, northern ireland and the isle of man (njrew) from united kingdom that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event Description
It was reported that, after undergoing a left thr revision surgery (b)(6) 2016 due to adverse soft tissue reaction to particulate debris, the patient experienced another adverse soft tissue reaction to particulate debris that made necessary a second revision surgery on (b)(6) 2020.During this procedure, all the femoral head, acetabular cup and acetabular liner were explanted.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a first thr revision surgery with a hip prosthesis construct that included a polarstem femoral prosthesis and that required a second revision surgery due to specific reasons.As such, no further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXINIUM FEM HD 12/14 36 MM M/+4
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16285898
MDR Text Key308664782
Report Number1020279-2023-00238
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010477286
UDI-Public03596010477286
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K022958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71343604
Device Catalogue Number71343604
Device Lot Number15LM03875
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/31/2023
Initial Date FDA Received02/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PN: 71331956 / LOT: 14CM12634.; PN: 71332756 / LOT: 14AM02161.; PN: 75100476 / LOT: B1514185.
Patient Outcome(s) Required Intervention; Other;
Patient Age65 YR
Patient SexFemale
-
-