ZIMMER BIOMET, INC. 36MM LINER IMPACTOR; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  malfunction  
Event Description
It was reported that the impactor broke.No additional information is available.
 
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following section was corrected: d4 (lot number).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h6, h10 proposed annex g code: mechanical (g04) - impactor reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information is available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified the post is fractured and missing.Wear & tear is seen that indicates repeated use.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM LINER IMPACTOR
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16286842
MDR Text Key308714086
Report Number0001822565-2023-00107
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00430902900
Device Lot Number61543997
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-