• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Erratic or Intermittent Display (1182)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  Injury  
Manufacturer Narrative
Initial reporter full hospital name: (b)(6).Health effect - clinical code - subcutaneous emphysema.The device referenced in this report was not returned to olympus for evaluation.The root cause cannot be determined at this time.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
A user facility reported to olympus that during a laparoscopic procedure with the high flow insufflation unit the display went out halfway through inflation.It continued to inflate but no information would show on the display; the error occurred intermittently.It was later stated that this had happened during several procedure causing ¿subcutaneous emphysema¿ in the abdominal area and some to the head (also stated as two incidences).A field service technician attended a later laparoscopic procedure and no anomaly was detected.The technician set the standard pressure values and the surgery was performed without complication.Later in conversation with the surgeon, the surgeon stated they also replaced the trocars with a different model.The extent of the emphysema was unable to be specified.It was stated intervention was needed but specific were unable to be provided.Additional information has been requested, if received, a follow up report will be sent.This issue involves two patient identifiers.Patient identifier (b)(6) is for the first instance of an emphysema occurrence.Patient identifier (b)(6) is for the second instance of an emphysema occurrence.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, it is very unlikely that an olympus device caused or contributed to the patient's development of emphysema.In addition, the report of no display/intermittent error could not be reproduced since the device was not returned for evaluation.Therefore, the root cause of both events could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16287699
MDR Text Key308706071
Report Number3002808148-2023-01022
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-