• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. OPTIPAC 80 BIOMET BONE CEMENT R; OPTIPAC (E) : BBCR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. OPTIPAC 80 BIOMET BONE CEMENT R; OPTIPAC (E) : BBCR Back to Search Results
Model Number N/A
Device Problems Chemical Problem (2893); Improper Chemical Reaction (2952)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source- japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that bone cement hardened faster than normal.It took about 5 minutes from the start of the mixing.Fifteen minute delay to replace the product.No health consequences to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The product was not returned or pictures were not provided.Visual device evaluation could not be performed.A retained sample of the batch has been tested in the laboratory under standardized conditions.No unusual behaviour during mixing, handling or setting.The reported behaviour of the cement cannot be reproduced.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.This device is used for treatment.The reported event is not related to a medical condition.A review of the medical record is not applicable.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC 80 BIOMET BONE CEMENT R
Type of Device
OPTIPAC (E) : BBCR
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16290058
MDR Text Key308824120
Report Number3006946279-2023-00011
Device Sequence Number1
Product Code MBB
UDI-Device Identifier04547038258066
UDI-Public(01)04547038258066(17)231130(10)AY49AB2114
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Model NumberN/A
Device Catalogue Number110035376
Device Lot NumberAY49AB2114
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/16/2023
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received02/14/2023
Supplement Dates FDA Received02/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-