• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-CABLE, MONOPOLAR; ELECTRICAL-ONLY MEDICAL DEVICE CONNECTION CABLE, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-CABLE, MONOPOLAR; ELECTRICAL-ONLY MEDICAL DEVICE CONNECTION CABLE, REUSABLE Back to Search Results
Model Number A0393
Device Problems Melted (1385); Smoking (1585); Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2023
Event Type  malfunction  
Manufacturer Narrative
The subject device was discarded, however, pictures of the device shows the cable below the protective boot cover at male connector side is damaged and completely separated.The investigation is still ongoing; however, if additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Event Description
The olympus sales representative reported the customer high-frequency monopolar cable had ¿sparked, melted and there was smoke¿.The customer reported event was found occurred during a transurethral resection of bladder tumor.The intended procedure was completed using another cable.There was a 5-minute delay and no additional anesthesia or sedation was required.The device was inspected prior to use but no anomalies were reported.No death or injury and no impact to patient or other has been reported to olympus.No device has been disposed by the customer following the procedure and will not be returned.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, it is likely the cable sparks/melts/smokes occurred due to wear and tear, in combination with improper handling.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-CABLE, MONOPOLAR
Type of Device
ELECTRICAL-ONLY MEDICAL DEVICE CONNECTION CABLE, REUSABLE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16294692
MDR Text Key308915282
Report Number9610773-2023-00415
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04042761001656
UDI-Public04042761001656
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944201
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA0393
Device Catalogue NumberA0393
Device Lot Number012999
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/05/2023
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received04/26/2023
Supplement Dates FDA Received05/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-