Model Number PM2152
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2023
Event Type  Injury  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported that the device was not pacing and was unable to be interrogated via inductive telemetry during an in clinic follow-up.The device was explanted and replaced to resolve the event and the patient was in stable condition.
 
Manufacturer Narrative
The reported events of loss of pacing and failure to interrogate were confirmed.The device was received with no telemetry communication and no output.A visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found at end of service level.A feed through leak test was performed, indicating a device hermeticity breach.This is consistent with feed through damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.The hybrid circuitry was tested by connecting to an external power source and the results indicated elevated current drain.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SBP DR INDUCTIVE PACEMAKER, JPD
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16295005
MDR Text Key308761298
Report Number2017865-2023-03027
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Model NumberPM2152
Device Lot NumberP000092493
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/16/2023
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received04/10/2023
Supplement Dates FDA Received04/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-