This is filed to report leak and air embolism, requiring intervention.It was reported that a mitraclip procedure was performed to treat mitral regurgitation (mr) grade 4 with a prolapsed posterior leaflet.Upon insertion of the steerable guide catheter (sgc) in the left atrium, an air bubble in the atrial auricle was visible.The sgc was removed and the air was aspirated.Another sgc was used to complete the procedure.Two clips were implanted with no reported issue, reducing mr to grade <1.There was no clinically significant delay in the procedure and no adverse patient sequelae.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported leak.The air embolism appears to be related to the leak.Air embolism is listed in the instructions for use as known a possible complication associated with mitraclip procedures.The reported unexpected medical intervention was a result of case-specific circumstance.There is no indication of a product issue with respect to manufacture, design, or labeling.
|