U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-180
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the reported issue was not confirmed.In addition, a worn-out scope socket/poor connection, faulty main printed circuit board, and air pump failure to work at high setting were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported the evis exera ii xenon light source brightness adjustment fails to work.As a result, it is difficult to make tissue distinguishable and the image is too bright.There was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the reported event could not be identified, nor could the phenomenon be confirmed/reproduced.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16309368
MDR Text Key309256761
Report Number3002808148-2023-01130
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170215544
UDI-Public04953170215544
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/13/2023
Initial Date FDA Received02/06/2023
Supplement Dates Manufacturer Received03/27/2023
Supplement Dates FDA Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-