(b)(4).The reported complaint of "high baseline alarm" was confirmed by the field service representative.As a result, the pump assembly was replaced.No part or recorder strip was returned to teleflex chelmsford for investigation.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the high baseline alarm.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.Other remarks: n/a.Corrected data: n/a.
|