Livanova deutschland manufactures the bubble sensor detector.The incident occurred in berlin, germany.A livanova field service representative was dispatched to the facility to investigate the device.Bubble sensor was checked, contact problems were eliminated.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
H10: through follow-up communication livanova learned the involved device serial number.Model and serial number have been added to the dedicated d.4 section.In addition, livanova learned that bubble sensor detector was triggering false alarm (the pump is stopped by the bubble alarm even if no air is present in the monitored line).The over-sensing issue has unlikely possibility to cause patient issue since the alarm can always be cleared/overridden by user, once the line is verified to be free of air, and the perfusion can be easily restarted.Therefore, for the above rationale the event has been reassessed as not reportable.It cannot be ruled out that the most likely root cause of the reported event is a temporary electrical failure, as a false contact, a bad/loose cable connection, or oxidation of electronics which was definitely solved during field service activity.
|