RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problem Operating System Becomes Nonfunctional (2996)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Manufacturer Narrative
The device was received by resmed and an evaluation was performed.The reported complaint could not be reproduced during evaluation and the device operated per specifications.The device was serviced, cleaned, calibrated and tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device touchscreen was intermittently unresponsive.The device was not in patient use when the reported event occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key16326557
MDR Text Key309371978
Report Number3004604967-2023-00095
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)220111(10)1583194
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2023
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/10/2023
Date Manufacturer Received01/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-