• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Defective Component (2292); Power Problem (3010)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event and model#: udi section are unknown, no information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the on/off switch was broken.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Device evaluation: one device was returned for investigation where it was found that customer complaint was confirmed.The pcb board and switch were not working leading to the device not having power.The connection between the the pcb and the power switch were damaged.The device is deemed beyond economical care and will be scrapped.A manufacturing device history review was not done as the device is beyond a year from the manufacturing date.Service history review identified this device had not been in for service previously.
 
Event Description
Additional information received via email on 8-june-2023.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16326764
MDR Text Key309079184
Report Number3012307300-2023-01040
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/26/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-