Model Number EL5ML
Device Problems Failure to Fire (2610); Mechanics Altered (2984); No Apparent Adverse Event (3189)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 01/11/2023
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic sigmoidectomy, the clip could not be fed into the jaw properly.The device was fired without clip fed properly, so the oozing occurred.The device was used on the blood vessels.The device was removed, and the clip came out when the device tip was shaken.Another device was used to complete the case.No patient consequences.
 
Manufacturer Narrative
(b)(4).Batch # x96225.Additional information was requested and the following was obtained: "semi-opened which it was planned to open it later, and sutured to hemostasis.The patient is stable after the operation." "did the patient need modified post op care due to the oozing? no.It is stated that the procedure was "semi-opened which it was planned to open it later".Was the procedure originally planned to be converted to open or was it converted to open because of the oozing that occurred? the procedure originally was planned to be converted to open." investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no damage to the external components.In addition, the tyvek was returned along with the instrument.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lockout after all the clips have been fired; therefore a potential cause for the customer reported experience is the firing of all of the clips, as a result, the instrument could no longer be fire due to the activation of the lockout mechanism.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.The instrument has an orange indicator that appears on the top of the handle as a reference for the user as to the number of clips remaining.The event described could not be confirmed as the device was returned empty.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: when the 13th clip is fired, an orange bar will begin to appear in the indicator window on top of the device handle.The orange bar fills the indicator window when the final clip is fired.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16327132
MDR Text Key309127838
Report Number3005075853-2023-00875
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX96225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Initial Date Manufacturer Received 01/13/2023
Initial Date FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-