CONMED UTICA VCARE 200A - MEDIUM; CANNULA, MANIPULATOR/INJECTOR, UTERINE Back to Search Results
Model Number 60-6085-201A
Device Problem Failure to Deflate (4060)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event Description
The customer reported that the device 60-6085-201a vcare 200a ¿ medium was being used on (b)(6) 2022 and the ¿balloon would not deflate when surgeon attempted to remove device from the uterus.Once removed balloon would still not deflate.A new device was opened and used.¿ there was no report of patient user impact, medical/surgical intervention or prolonged hospitalization.The procedure was completed with an alternate same device.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Event Description
The customer reported that the device 60-6085-201a vcare 200a ¿ medium was being used on 2dec22 and the ¿balloon would not deflate when surgeon attempted to remove device from the uterus.Once removed balloon would still not deflate.A new device was opened and used.¿ there was no report of patient user impact, medical/surgical intervention or prolonged hospitalization.The procedure was completed with an alternate same device.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Manufacturer Narrative
The device will not be returned, and no photographic evidence was provided.Therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 12 reports, regarding 15 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following: to remove vcare from its sterile packaging and inspect for damage caused by shipping.Discard if any damage is noted.Draw 7-10 cc of air into a standard syringe and insert it into the luer connection at the end of the pilot balloon.Test the intrauterine balloon by injecting air with the syringe and checking to see if balloon remains inflated.If balloon does not remain inflated, do not use.Discard and select another vcare unit.After the successful balloon test, deflate the balloon by removing all air with the syringe.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCARE 200A - MEDIUM
Type of Device
CANNULA, MANIPULATOR/INJECTOR, UTERINE
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key16340812
MDR Text Key309371007
Report Number1320894-2023-00033
Device Sequence Number1
Product Code LKF
UDI-Device Identifier10653405003793
UDI-Public(01)10653405003793(17)240925(10)202209261
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-6085-201A
Device Catalogue Number60-6085-201A
Device Lot Number202209261
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/17/2023
Initial Date FDA Received02/09/2023
Supplement Dates Manufacturer Received03/27/2023
Supplement Dates FDA Received03/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2022
Is the Device Single Use? Yes
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-