Model Number 6662 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 01/16/2023 |
Event Type
Injury
|
Event Description
|
Related manufacturer reference number:1627487-2023-00365.It was reported that the patient experienced side effects of speech arrest when reaching with their left arm, light-headed, lessening of tremor benefits.Reprogramming was unable to resolve the issue.As a result, surgical intervention was undertaken on (b)(6) 2023 wherein the right lead was explanted to address the issue.During the procedure, the ipg became unable to communicate with external devices.The device was confirmed to have been placed into surgery mode.Troubleshooting was unsuccessful.As a result, surgical intervention may be undertaken in the future.
|
|
Manufacturer Narrative
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Additional information indicates surgical intervention was undertaken on (b)(6) 2023 wherein a right lead was implanted and the ipg and both extensions were explanted and replaced to address the issue.
|
|
Manufacturer Narrative
|
The report of inoperable ipg was confirmed.Analysis of the returned ipg found the device was set to surgery mode while the patient underwent lead revision surgery.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Analysis of the device found that the system is in service app mode and could not be recovered.
|
|
Search Alerts/Recalls
|