Model Number HL-390
Device Problem Leak/Splash (1354)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the fluid warmer was leaking from the front.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Date of event and udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H6: event methods, evaluation, and conclusion codes: updated.One warmer device was received for investigation.During visual inspection, the warmer was observed to have a cracked tank cover.The reported issue was confirmed during functional testing when water was observed to leak from the device.The leakage was traced to cracks on the water tank, around the screws that fasten its cover.The cracking was attributed to overtightening of the screws.As no manufacturing related root cause could be established, no review of device history records was conducted.Review of service records indicate the device was not previously serviced.Due to the age of the device, the device was deemed beyond economical repair and will be decommissioned.This issue was escalated to an internal scar and occurrence is being monitored via trend analysis.If the occurrence of this issue increases significantly, additional corrective actions will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16351774
MDR Text Key309297712
Report Number3012307300-2023-01127
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/18/2023
Initial Date FDA Received02/10/2023
Supplement Dates Manufacturer Received02/15/2023
Supplement Dates FDA Received03/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-