Model Number HL-90
Device Problem Failure to Calibrate (2440)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event and d4:udi section are unknown, no information has been provided to date.The device was received and inspected.Investigation confirmed the calibration issue and traced it to an outdated and faulty pcb (printed circuit board) unit.A dhr (device history review) review determined the device was manufactured in 2005 and has not received service since that date.Issue is attributed to age and condition of the device.No action taken due to the age and condition of the device.It was deemed beyond economical repair and it will be scrapped.
 
Event Description
It was reported the warmer will not calibrate.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16352315
MDR Text Key309379781
Report Number3012307300-2023-01134
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2023
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-