• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGFA NV PORTABLE RADIOGRAPHY UNIT; FILM, RADIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGFA NV PORTABLE RADIOGRAPHY UNIT; FILM, RADIOGRAPHIC Back to Search Results
Model Number DX-D100
Device Problems Detachment of Device or Device Component (2907); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2022
Event Type  malfunction  
Event Description
During a routine patient imaging procedure using an agfa dx-d 100 portable radiographic unit, the counter weight cable snapped and the horizontal arm with attached x-ray tube head fell toward the patient bed.The x-ray head collimator landed on the bed railing and that stopped the fall.Neither patient nor personnel were harmed by the cable failure.After the incident our engineer inspected this agfa dx-d 100 and determined that following the counter weight cable failure, tension was not released from the cable attached to the backup safety latch.Since tension was not released, the backup safety latch did not engage the vertical arm and stop the x-ray head from falling toward the bed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTABLE RADIOGRAPHY UNIT
Type of Device
FILM, RADIOGRAPHIC
Manufacturer (Section D)
AGFA NV
mortsel
BE 
MDR Report Key16352924
MDR Text Key309405904
Report NumberMW5114843
Device Sequence Number1
Product Code IWZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDX-D100
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/09/2023
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-