It was reported that during a device changeout procedure from another manufacturer's biventricular (biv) implantable cardioverter defibrillator (icd) to this cardiac resynchronization therapy defibrillator (crt-d), left ventricular (lv) pacing inhibition was observed for an unspecified duration once the chronic lv lead was connected to the new device.It was noted that the lv was programmed from lv1>rv on the previous device, and was then programmed to 3>4 for pacing on the crt-d.It was determined that the crt-d was not placed in electrocautery mode, resulting in oversensing and pacing inhibition due to electromagnetic interference (emi) from the procedure.The patient was pacer dependent with no escape rhythm.The lead configuration was confirmed as appropriate, and the procedure was completed successfully.This crt-d system remains in service, and the patient will continue to be monitored remotely.No adverse patient effects were reported.
|