BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 3822
Device Problems Material Rupture (1546); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).
 
Event Description
It was reported that balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and severely calcified right coronary artery.A 10/3.00 flextome cutting balloon was selected for use.During the procedure, the balloon ruptured at 8 atmospheres within 5 seconds and a pinhole was noted.The balloon was removed normally without any additional intervention and the procedure was completed with another of the same device.No complications were reported, and the patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME CUTTING BALLOON
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16355299
MDR Text Key309381295
Report Number2124215-2023-04729
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/27/2023
Device Model Number3822
Device Catalogue Number3822
Device Lot Number0025925590
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2023
Initial Date FDA Received02/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight60 KG
-
-