MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37601
Device Problems Loss of Data (2903); Data Problem (3196)
Patient Problem Shaking/Tremors (2515)
Event Date 01/14/2023
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id : 8840, serial#: unknown, product type: multiple therapy product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient had a programming session yesterday with an n'vision clinician programmer.Caller reports she did not know the clinic still had that programmer but the n'vision programmer was used instead of the clinician programmer.Caller reports patient started to have his tremor returned this morning and when he checked with his patient programmer, error code with a doctor icon 574 occurred.Reviewed error code 574.Additional information was received from the manufacturer representative (rep).The cause was unknown.Actions taken to resolve the issue were the ins was reprogrammed.This information was confirmed with the physician/patient.(b)(6) 2023 initial reporter health professional (rep): additional information received from the manufacturer¿s representative (rep) reported the error code was resolved.Additional information received from the manufacturer¿s representative (rep) reported the patient¿s settings were lost after they used the n¿vision programmer as the physician used the clinician programmer to interrogate the neurostimulator and it cleared the settings and reset it.The rep met with the patient at a later date and interrogated the neurostimulator with the tablet and reprogrammed it back to the original settings.The patient went home with therapeutic delivery.The rep removed the clinician programmers from the office to prevent future incidents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16361244
MDR Text Key309370290
Report Number3004209178-2023-01958
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00763000175719
UDI-Public00763000175719
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2020
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Date Manufacturer Received02/08/2023
Date Device Manufactured02/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-