MAUDE Adverse Event Report: MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number ONYXNG27515X
Device Problems Deflation Problem (1149); Difficult to Remove (1528); Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  malfunction  
Event Description
An attempt was made to use one onyx frontier coronary drug eluting stent to treat a non-calcified, non-fibrotic lesion.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop/tray.The stent did not have anything damaged once removed from the package.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.The stent passed smoothly through the lesion.It was reported that balloon deflation difficulties occurred, and the device would not deflate at the lesion site.The stent was inflated at nominal pressure of 13 atm.When checking the stent with rx it was discovered that the stent was broke.An attempt was then made to remove the stent but the device was difficult to deflate.Pressure had to be applied to retract the stent.There was no major consequences for the patient.The patient was reported to be alive with no injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis the device was received for analysis.Only the stent delivery system returned for analysis.No stent returned with the stent delivery system.Numerous kinks were evident along the hypotube.It was not possible to perform negative prep due to the condition of the returned device.The transition shaft appeared stretched and kinked when compared to an in-house onyx frontier.The entry port was torsionally kinked and deformed.Blood was visible in the inflation lumen.Kinks and stretching were evident along the distal shaft.Kinks were evident on the inner member.Bunching was evident on the inner member at the proximal balloon bond/inflation lumen.Kinks were evident to the inner member and outer shaft between the markerbands.The balloon was deflated on device return.The balloon material appears to have folded distally over the distal marker band on retraction from the patient.It was not possible to preform deflation testing due to the condition of the proximal bond/ inflation lumen.No other deformation evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: annex d code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16361352
MDR Text Key309602465
Report Number9612164-2023-00618
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberONYXNG27515X
Device Catalogue NumberONYXNG27515X
Device Lot Number0011388814
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-