Model Number 106 |
Device Problems
No Apparent Adverse Event (3189); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
|
Event Description
|
It was reported in clinic notes that there was concern for a vns device malfunction with the battery level decreasing.The battery status is not low.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
|
|
Event Description
|
It was later reported that the patient is not moving forward with a replacement as the device has plenty of battery life left with low settings.M103-m1000 generators check impedance, battery, and output current every 24 hours when programmed on and will display a warning if any abnormalities were detected during that 24 hour check.The device was checked during the patient's surgical consult.Because there is no indication of high impedance, low output current, etc., it can be reasonably concluded that the vns is functioning within normal limits.No other relevant information has been received to date.
|
|
Event Description
|
Information was later received that the patient had a prophylactic battery replacement.The explanted device is not available for return to the manufacturer, indicating that it has likely been discarded.
|
|