Model Number SUPPORT ARM 176 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/20/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the support arm holding the patient tubes broke.There was no patient harm.Manufacturer's ref #: (b)(4).
|
|
Manufacturer Narrative
|
The support arm was repaired by our field service engineer and was cleared for clinical use.The support arm has been successfully tested for mechanical strength and is designed according to standard.The root cause of the broken support arm has not been conclusively determined.However, the breaking point indicates that it most likely has been exposed to a mechanical force greater than it is designed to sustain.
|
|
Event Description
|
Manufacturer's ref #: (b)(4).
|
|