MAUDE Adverse Event Report: EDWARDS LIFESCIENCE SWAN GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCE SWAN GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.Further evaluation regarding related quality issues is under investigation.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that a swan ganz was unable to pace from the beginning of use after the catheter was inserted and connected to an external pacemaker.The catheter was intended to be used for tavi procedure.The issue was resolved by replacing the catheter and the procedure was performed successfully.It is unknown if the patient had cardiac conduction detect.Date of event is also unknown.Patient demographic information was requested but unavailable.The device was discarded at the hospital.There were no patient complications reported.
 
Manufacturer Narrative
Engineering evaluation was completed and concluded that a product risk assessment escalation was met earlier.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCE
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key16375978
MDR Text Key309563801
Report Number2015691-2023-10804
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPE074F5
Device Lot Number64496181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received02/14/2023
Supplement Dates Manufacturer Received03/21/2023
Supplement Dates FDA Received03/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-