Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hemorrhage/Bleeding (1888); Loss of consciousness (2418); Convulsion/Seizure (4406)
|
Event Date 02/09/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date of event is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A webform complaint was received in which it was reported a customer received a "glucose reading unavailable" message on the abbott diabetes care (adc) device and was unable to obtain readings.As a result, customer experienced a loss of consciousness and/or seizures, "blood leaking from sensor hole", and did not report any third-party treatment.No further details were provided and attempts to gather additional information were unsuccessful.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is properly seated and no physical damage was observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.No malfunction or product deficiency was identified.Therefore, the issue is not confirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A webform complaint was received in which it was reported a customer received a "glucose reading unavailable" message on the abbott diabetes care (adc) device and was unable to obtain readings.As a result, customer experienced a loss of consciousness and/or seizures, "blood leaking from sensor hole", and did not report any third-party treatment.No further details were provided and attempts to gather additional information were unsuccessful.There was no report of death or permanent impairment associated with this event.
|
|