Catalog Number 292.620 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/23/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial , a follow-up will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in columbia as follows: it was reported that on (b)(6) 2023, during the medial malleolus osteosynthesis surgery, the 1.25 pin was passed that was necessary to put 4.0 cannulated screws, at the time of brocade the bit was mispositioned which caused the fracture of the aforementioned pin.This report is for one (1) guidewire ø1.25 w/thread-tip w/trocar l1.This is report 1 of 2 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D7a.Is this a single-use device? added.G4 (510k): unknown; device is not distributed in the united states, but is similar to device marketed in the usa.H4: device manufacturer date added.Part# 292.620.Lot # 3l04909.Manufacturing site: werk balsthal.Release to warehouse date: 10 jan 2019.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|