Model Number CDVRA500Q
Device Problem Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Event Description
It was reported by the patient that the implantable cardioverter defibrillator was unable to pair with mobile application via bluetooth.No intervention was performed.There were no patient consequences.
 
Event Description
New information received notes that the implantable cardioverter defibrillator (icd) was in bluetooth telemetry lockout.The icd was interrogated, and the bluetooth telemetry was restored.No patient consequences were reported.
 
Manufacturer Narrative
The reported event of ble unavailable was resolved by interrogating the device using a merlin programmer, this is indicative of a ble lockout which is per device design.No anomalies found and no further investigation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16383844
MDR Text Key309635260
Report Number2017865-2023-10074
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000080166
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received02/15/2023
Supplement Dates Manufacturer Received04/06/2023
06/02/2023
Supplement Dates FDA Received04/26/2023
06/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA
Patient Age55 YR
Patient SexMale
-
-