Model Number ALLURA XPER FD10
Device Problems Positioning Failure (1158); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the angle of the arm was unable to be changed.No harm has been reported to philips.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was in clinical use when the issue occurred.The customer reported that the system had issues with the changing angles during procedures and gets stuck when trying to change.The philips field service engineer (fse) inspected the system onsite and could not replicate the reported issue.The reported problem did not reoccur.Fse performed a current calibration on stand movements as this pertains to the issue reported and the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
boschdijk 525
postbus 90050
eindhoven 5600 -PB
NL   5600 PB
MDR Report Key16384280
MDR Text Key309641552
Report Number3003768277-2023-00869
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00884838059030
UDI-Public00884838059030
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722010
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2023
Initial Date FDA Received02/16/2023
Supplement Dates Manufacturer Received02/16/2023
Supplement Dates FDA Received11/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-