• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0635
Device Problem Failure to Transmit Record (1521)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  malfunction  
Event Description
According to the reporter, the capsule did not register the exam.The capsule did not transmit data to the recorder 24 hours after the placement to the patient esophagus.There was no data present on the recorder.The patient undergo sedation during the initial procedure and the repeat procedure was needed but not scheduled yet.There was no patient or user harm.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS  20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
IS   20692
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16387679
MDR Text Key309898006
Report Number9710107-2023-00027
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101369707
UDI-Public07290101369707
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/07/2024
Device Model NumberFGS-0635
Device Catalogue NumberFGS-0635
Device Lot Number57566F
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/09/2023
Initial Date FDA Received02/16/2023
Date Device Manufactured08/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-