• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION Back to Search Results
Model Number 8120
Device Problem Failure to Analyze Signal (1539)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2023
Event Type  malfunction  
Manufacturer Narrative
A device history record review is performed on each device reported in a mdr reportable event along with other methods of investigation of this mdr report.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
 
Event Description
It was reported that the device had failed accuracy test, failed the plunger force accuracy test and tried to calibrate using the software and it failed, giving message that the unit needs to be sent in for repair.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key16390789
MDR Text Key309746945
Report Number2016493-2023-113795
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10885403812002
UDI-Public(01)10885403812002
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8120
Device Catalogue Number8120 ALARIS PCA MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2023
Initial Date FDA Received02/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-