Lot Number COV2010029
Device Problem No Device Output (1435)
Patient Problem Insufficient Information (4580)
Event Date 01/25/2023
Event Type  malfunction  
Event Description
Invalid result (s).Called in because she purchased a flowflex kit and no results displayed.No c or t line appeared.She followed the directions exactly and dispensed the 4 drops into the s well.The desiccant pack was inside the test cassette pouch.The kit was purchased at kroger.
 
Event Description
Invalid result (s).Called in because she purchased a flowflex kit and no results displayed.No c or t line appeared.She followed the directions exactly and dispensed the 4 drops into the s well.The desiccant pack was inside the test cassette pouch.The kit was purchased at kroger.
 
Manufacturer Narrative
Batch records, final product manufactured and qc records were reviewed for lot cov2010029.No abnormal issue was found in the manufacturing process, technical testing and quality control inspection, the manufacturing process complied with the dmr.Test results of retention samples meet with the qc criteria.The reported issue was not found.This complaint is not verified.In this follow-up report, the following information has been updated from the initial report upon completion of the internal investigation.The following fields are updated: b4, "date of this report" - date of follow-up report.G6, "type of report" - follow-up #1.H2, "if follow-up, what type?" - updated to "additional information" and "device evaluation" h6 "adverse event problem" - event problem and evaluation codes such as "investigation findings" and "investigation.Conclusions" are added per investigation.H10 "additional narrative/data" - added manufacturer's narrative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWFLEX COVID-19 ANTIGEN HOME TEST
Type of Device
ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP
Manufacturer (Section D)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer (Section G)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer Contact
qiyi xie
5850 oberlin drive #340
san diego, CA 92121
8588758011
MDR Report Key16392393
MDR Text Key309767720
Report Number2531491-2023-00490
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
EUA210494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Lot NumberCOV2010029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/25/2023
Initial Date FDA Received02/16/2023
Supplement Dates Manufacturer Received01/25/2023
Supplement Dates FDA Received03/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-