• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71857-01
Device Problems Device Alarm System (1012); No Audible Alarm (1019); Application Program Problem (2880)
Patient Problems Hypoglycemia (1912); Blurred Vision (2137); Dizziness (2194)
Event Date 02/02/2023
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensors and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
An alarm issue was reported with the adc application in use with an android phone (samsung sm-a528b) with operating system v.13 (full software v.2.8.1.9305).Customer experienced a signal loss and was unable to receive glucose alarms.As a result, customer was not alerted of changes in glucose level and experienced dizziness, blurred vision and was unable to self-treat, requiring treatment of juice by third-party.There was no report of death or permanent injury associated with this event.
 
Event Description
An alarm issue was reported with the adc application in use with an android phone (samsung sm-a528b v.2.8.1.9305) with with android 13 operating system.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the customer experienced dizziness, blurred vision and was unable to self-treat, requiring treatment of juice by third-party.There was no report of death or permanent injury associated with this event.The impacted product associated with this complaint is on market in the united states as well as the following countries ous: au, at, be, hr, cz, dk, fi, fr, de, gr, il, it, kw, lu, nl, nz, no, pl, pt, qa, sa, es, se, ch, tw, ae, gb, ca.Field action fa1010-2023 was issued to all impacted countries 08feb23.
 
Manufacturer Narrative
Adc has identified a software defect for the freestyle libre 2 application with the android 13 os where the application may experience intermittent signal loss.As a result, the application user may not receive glucose results and/or glucose alarms and may not be alerted of low or high glucose conditions.Based on the investigation, this complaint is confirmed.This issue was addressed in the field by adc fa1010-2023.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key16397100
MDR Text Key309756135
Report Number2954323-2023-07262
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K210943
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/30/2023
Device Model Number71857-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/03/2023
Initial Date FDA Received02/17/2023
Supplement Dates Manufacturer Received04/19/2023
Supplement Dates FDA Received05/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberQUALITY RECORD QR869957
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-