Model Number D134805
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2022
Event Type  malfunction  
Event Description
A patient underwent an ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter for which biosense webster¿s product analysis lab identified that the pebax was broken and an electrode was bent with no parts exposed.The finding was identified on (b)(6) 2023.During the procedure, it was reported that there were noises on the proximal electrodes.Surgery was delayed due to the reported event for 15 minutes.Action taken when event occurred was nothing because it was the second catheter for this procedure.Procedure was successfully completed.No patient consequences.A brit on the proximal electrodes was detected; a second device had to be used.The doctor did not notice any odd condition of the catheter or electrode when the problem occurred.Bad/partial ecg (body surface or intracardiac) is not mdr-reportable.Broken pebax is mdr-reportable.
 
Manufacturer Narrative
The product investigation was completed.Device evaluation details: visual analysis revealed a that the pebax was broken and an electrode bent with no parts exposed.An electrical test was performed, and no electrical issues were found.A manufacturing record evaluation was performed for the finished device (b)(4) number, and no internal actions related to the reported complaint condition were identified.The electrical issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain the following recommendation: ecg noise is typically generated as the result of the improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify the proper connection.It is recommended to turn off the notch filter for this verification.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THMCL SMTCH SF BID, TC, D-F
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16397973
MDR Text Key310168767
Report Number2029046-2023-00324
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835010183
UDI-Public10846835010183
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD134805
Device Catalogue NumberD134805
Device Lot Number30875187L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2022
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-