Model Number 777000
Device Problem Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus, and it was observed that the output is too low.This complaint is most likely the result of a faulty power supply unit.Testing and inspection also found: g600 error code, this error occurs if the auxiliary board does not respond on time.This complaint is most likely the result of a faulty auxiliary board.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
A user facility returned the olympus asset, g400 generator-gyrus, to the olympus service center.Upon inspection and testing of the returned device, it was observed that the output is too low and fails.This report is being submitted for the event found during evaluation.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see updates to h4, h6 and h10.The device history record was not reviewed as the device has been previously repaired by the olympus service team.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it¿s likely the output was too low because of a faulty power supply unit.The final root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G400 GENERATOR, GYRUS
Type of Device
ELECTROSURGICAL SYSTEM
Manufacturer (Section D)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer (Section G)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16406075
MDR Text Key309989016
Report Number3003790304-2023-00083
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009004
UDI-Public00821925009004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050550
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number777000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2022
Was the Report Sent to FDA? No
Date Manufacturer Received05/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-