Reportable based on device analysis completed on 07feb2023.It was reported that a device kinked occurred.The target lesion was located in the left anterior descending artery.A 10mmx3.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the device was kinked.The device was removed with the normal method and the procedure was completed with another of the same device.No patient complications reported.However, device investigations revealed a perforation in the outer extrusion at 5.6cm proximal from the tip.
|
Device evaluated by manufacturer: a wolverine cb 10mmx3.50mm, catheter was returned for analysis.Based on the potential hazards/failure modes identified and the complaint report, the following attributes were examined: a visual examination identified that the balloon was folded.A microscopic examination of the balloon material and marker bands identified no issues which could potentially have contributed to this complaint.All blades were intact within their pads and fully bonded to the balloon material.A visual and tactile examination found a kink in the hypotube.The kink was located at 65.5cm distal from the strain relief.An examination of the distal extrusion identified that the distal extrusion was kinked at 9cm; 9.4cm and 10cm distal from the port.A microscopic examination of the inner/wire lumen noted bunching of the extrusion at approximately 5.2cm proximal from the tip.Further examinations noted a perforation in the outer extrusion at 5.6cm proximal from the tip.It is likely that as a result of the bunching in the wire lumen, when an attempt was made by the physician to pass the wire through the bunched lumen, it resulted in the guidewire puncturing out through the inner and outer lumen.The marker bands and tip and of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.
|