Calibration signals were slightly lower than expected.Qc results fluctuated.No further information was provided by the customer.Due to the limited information provided, the cause of the event could not be determined.Product labeling states: "the measured ca 15 3 value of a patient's sample can vary depending on the testing procedure used.The laboratory finding must therefore always contain a statement on the ca 15 3 assay method used.Ca 15 3 values determined on patient samples by different testing procedures cannot be directly compared with one another and could be the cause of erroneous medical interpretations.If there is a change in the ca 15 3 assay procedure used while monitoring therapy, then the ca 15 3 values obtained upon changing over to the new procedure must be confirmed by parallel measurements with both methods." the investigation did not identify a product problem.
|