Model Number 0998-00-0800-55 |
Device Problem
Component Incompatible (1108)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/13/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit has a bleed purge problem.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that before patient use, the cardiosave intra-aortic balloon pump (iabp) unit has a bleed purge problem.There was no patient harm reported.
|
|
Event Description
|
It was reported that during preparation, before patient use, the cardiosave intra-aortic balloon pump (iabp) unit had a bleed purge problem.It was just before the ball was to be put in place that they had a bleeding problem.There was no patient harm reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse noticed that there was water between the safety disk and the pim, so the fse did a water purge and the test passed.The fse did not see any problem when checking the machine.The fse found a problem with the printer.Estimate in progress.The fse changed the exec card because it had not be changed last year during preventive maintenance.Checking of the device according to the manufacturer's procedure.If information is received for the printer, we will reopen and update the complaint.
|
|
Event Description
|
N/a.
|
|