• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX SANTA ROSA OVATION IX; ILIAC LIMB Back to Search Results
Model Number TV-AB2980-N
Device Problems Deformation Due to Compressive Stress (2889); Unintended Movement (3026)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 02/08/2023
Event Type  Injury  
Event Description
The patient was treated for an abdominal aortic aneurysm (aaa) with the implant of the alto abdominal stent graft system.Sixteen (16) days post-procedure, it was identified that the right ovation ix iliac limb had moved down and thrombosed.There was bad out flow and a kink in the alto graft in the common iliac.Reintervention was attempted on (b)(6) 2023; however, the physician was unable to pass anything through the thrombosed system.A fem-fem bypass was reportedly planned, though it is unknown if it was completed.Patient status was not reported.
 
Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.
 
Manufacturer Narrative
The reported incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the incident could not be completed.No medical records nor medical imaging relevant to the reported incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging device, use, procedure, and/or anatomy relatedness to this incident could not be evaluated.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Type of Device
ILIAC LIMB
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
Manufacturer (Section G)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
,
santa rosa CA 95403
Manufacturer Contact
gary kirchgater
3910 brickway blvd
,
santa rosa, CA 95403
8009832284
MDR Report Key16418575
MDR Text Key309984448
Report Number3008011247-2023-00024
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTV-AB2980-N
Device Catalogue NumberTV-IL1414140-J
Device Lot NumberFS070722-64
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/08/2023
Initial Date FDA Received02/22/2023
Supplement Dates Manufacturer Received02/08/2023
Supplement Dates FDA Received04/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ALTO MAIN BODY (LN FS070722-64); OVATION IX ILIAC LIMB (LN FS092722-31); OVATION PRIME FILL POLYMER (LN FF071122-01)
Patient Outcome(s) Other;
Patient Age65 YR
-
-