H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.Sr8 is having issues holding a charge as the unit shuts off shortly after ac adapter is disconnected¿ is confirmed.The unit was charged at 100% and within 10 minutes or so it had already dropped down to 88%.The lithium battery needs to be replaced.The reported issue is due to an internal failure within the battery.H3 other text: evaluation findings are in section h11.
|