• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problems Pumping Stopped (1503); Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that the on/off push button of a s5 roller pump was defective since it tended to come loose.The issue occurred during maintenance activity.There was no patient involvement.
 
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in (b)(6) italy.Through follow-up communication livanova learned that the customer turns off the s5 system from the button present in the the ep (electronics and power) pack, instead of switch off the pump modules individually at the end of the procedure.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, on/off push button was replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
The affected roller pump was manufactured in 2016 and according to the complaints database review, no further issues have been submitted for this unit.The push button switch is a small, sealed mechanism that completes an electric circuit when it is pressed.When it is in the "on" position, a small metal spring inside makes contact with two wires, allowing electricity to flow.When it's off, the spring retracts, contact is interrupted, and current does not flow.If the switch does not properly latch, most likely is its internal spring to be faulty.Based on the above facts, it cannot be ruled out that a defective spring caused the reported malfunction.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See intial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key16427044
MDR Text Key310272338
Report Number9611109-2023-00080
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900405
UDI-Public(01)04033817900405(11)161208
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2023
Initial Date FDA Received02/23/2023
Supplement Dates Manufacturer Received03/31/2023
Supplement Dates FDA Received04/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-