Model Number 006173P
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problems Exposure to Body Fluids (1745); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2021
Event Type  Injury  
Event Description
It was reported that the temporary pacing electrode catheter was kinked.The patient was exposed to hazardous, blood, or bodily fluids.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the temporary pacing electrode catheter was kinked.The patient was exposed to hazardous, blood, or bodily fluids.
 
Event Description
It was reported that the temporary pacing electrode catheter was kinked.The patient was exposed to hazardous, blood, or bodily fluids.It was unknown what medical intervention was provided.
 
Manufacturer Narrative
The reported event is inconclusive because the complaint sample was not returned for evaluation and further investigation was not conclusive.Though a specific cause cannot be determined, a potential root cause for this event could be, "pulling distal wire to taut at mp5898 (assemble electrodes proximal/distal)".The device was used for treatment purposes.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "excessive bending, torquing, or kinking of the electrode catheter may cause damage to the catheter including damage to internal wires." "inspect the sterile package carefully for damage during transit or storage.Do not use the catheter if the package is damaged." "visually inspect the catheter, under sterile conditions, for kinks in the catheter shaft, integrity of the connector, condition of electrodes, and any other damage." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text: the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPORARY PACING ELECTRODE CATHETER
Type of Device
TEMPORARY PACING ELECTRODE CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16428102
MDR Text Key310181221
Report Number1018233-2023-01140
Device Sequence Number1
Product Code LDF
UDI-Device Identifier00801741011207
UDI-Public(01)00801741011207
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K800298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model Number006173P
Device Catalogue Number006173P
Device Lot NumberGFDX4510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-