Lot Number 0028492645 |
Device Problem
Entrapment of Device (1212)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that catheter entrapment on guidewire occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified right coronary artery.A 4.00 x 32mm synergy drug-eluting stent was advanced for treatment.However, during the procedure, the stent became stuck with a non-boston scientific guidewire.The devices were removed as one unit and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: synergy xd mr ous 4.00 x 32mm was returned for analysis.Returned device analysis found that the distal extrusion (including the wire lumen) was stretched and bunched.As a result of this damage, it was not possible to pass a recommended 0.014 inch guidewire through the stretched and bunched lumen.Further analysis identified that the hypotube was kinked at various locations along its length.
|
|
Event Description
|
It was reported that catheter entrapment on guidewire occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified right coronary artery.A 4.00 x 32mm synergy drug-eluting stent was advanced for treatment.However, during the procedure, the stent became stuck with a non-boston scientific guidewire.The devices were removed as one unit and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
|
Search Alerts/Recalls
|